Investigation of Quality Control Measurements with Aging X-Ray Machines

Agaku, M. R. and Akaagerger, N. B. and Ode, O. S. (2021) Investigation of Quality Control Measurements with Aging X-Ray Machines. Current Journal of Applied Science and Technology, 40 (13). pp. 16-21. ISSN 2457-1024

[thumbnail of 3545-Article Text-6579-1-10-20220914.pdf] Text
3545-Article Text-6579-1-10-20220914.pdf - Published Version

Download (372kB)

Abstract

The investigation of Quality Control measurement (QC) with ageing x-ray machines was undertaken. Ten x-ray machines were examined across different hospitals within Makurdi metropolis, Benue State using Gammex 330 kV meter and ten (10) X-ray Machines in government hospitals (GH) and private hospitals (PH) with age’s (4-12) years for new X-ray Machines and (22-47) years for aged X-ray Machines. With these, on the quality of diagnostic X-rays and patient’s dose were examined separately when a graph of age of the machine was plotted against the power n of the kVp and a graph of machines age was plotted against radiation intensity. Based on the analysis obtained in this study, the value of n for the new machines with ages (5-12) years has a value of and aged machines with ages (22-47) years has a value of. This shows that there is a relationship between age and n. In conclusion, the findings of the present study showed that the output radiation dose of new and aged x-ray units differs at the same kVp and mAs measurements, indicating that the x-ray tube deteriorates as its ages increase, thereby posing higher health risk to patients and as such, the study recommends that QC measurement be undertaken on x-ray machines at definite stages of the machines life to verify the effects of aging of a machine on patient's dose.

Item Type: Article
Subjects: Science Repository > Multidisciplinary
Depositing User: Managing Editor
Date Deposited: 15 Apr 2023 07:11
Last Modified: 25 May 2024 07:30
URI: http://research.manuscritpub.com/id/eprint/1609

Actions (login required)

View Item
View Item